By default the scale only sends weigh weight data to the POS when the POS requests it. This will tell you how to configure the scale to send the weigh weight data to the POS as soon as the scale has "stabilized". Follow the instructions and images below.


Go in to SETUP mode by pressing and holding the HOLD and ZERO buttons for about three seconds. You should see CONFG on the screen.

Image Added


Press the UNIT button until the screen changes to USEr. Press TARE to continue.

Image Added


Press the UNIT button until you see the Con1 screen. Press TARE to continue.

Image Added


Press the UNIT button until you see the oUtnd screen. Press TARE to continue.

Image Added


Press the UNIT button until you see the StAbl screen. Press TARE to continue.

Image Added


Press the ZERO three times to back out, at the Exit screen press TARE.

Image Added